All test sieves supplied by Endecotts are measured and inspected, in line with the specified standard, using sophisticated measuring equipment.
All test sieves are inspected to fully comply with either:
All Endecotts measurement apparatus is calibrated in line with specified arrangements which form part of our ISO 9001 Quality Management System – Registered with BSI, certificate No FM 24761. The calibration of our optical measuring equipment for mesh <= 10mm is carried out using glass objects which are calibrated by the ‘Physikalisch-Technische-Bundesanstalt’ (PTB), Braunschweig, Germany. The calibration of our measuring equipment for mesh > 10mm is traceable to a British calibration laboratory. The traceability to national and international standards is ensured
Compliant sieves are certified in line with ISO 3310-1 & 2 requirements.
All Endecotts test sieves are issued with a Certificate of Compliance with a unique, traceable, serial number.
Higher levels of certification, providing varying levels of measurement data, as detailed in clause 5.3.2 of ISO3310 – 1 & 2 may be specified by customers
Regularly used test sieves will eventually suffer from 'wear and tear' – which can mean inaccurate results. Sieves should therefore be re-inspected at appropriate intervals to ensure they continue to comply with their intended specification. In fact, regular Re-certification is essential if test sieves are to meet the requirements of the test sieve specification.